Nouveau processeur de poids HI6501 modernisé
En savoir plus !The HI 2151/30WC and HI 2110WI units will display a minus sign on the left hand side of the display to indicate the weight is a negative value. The weight controllers will display a weight change as long as the millivolt reading is positive. The HI 2151/30WC and HI 2110WI do not process negative millivolt readings lower than -0.5mV for weight signal input.
On the newer HI 2151/30WC controllers the word LOW is displayed when the -0.5mV range is exceeded. Locking on a particular negative weight did not convey the weight was now out of range in a negative value. This LOW display is seen when the load cell is damaged or they are lifting up on a load cell.