Sie haben keine Artikel in Ihrem Warenkorb.

Intelligenter wiegen mit der neuen HI5034-WS

Erfahren Sie mehr!
Vertrieb: 858-278-2900
Support: 858-292-2710
Besuchen Sie hardysolutions.cn

Why does my HI 1746WS read 50 lbs after doing a C2® calibration?

Last updated 10.02.2014 10:07 AM

Question

The C2® calibration on my HI 1746WS yields a weight of 50 lbs when I push the "Button". The scale is empty, so it should read 0 lbs. What should I change on my HI 1746WS to get an accurate calibration?

Answer

The reference point the HI 1746WS module uses for the "Button" calibration is the CalLo Weight. You should verify this is set to the amount of weight on the scale (zero if empty scale) at the time of the "Button" calibration.